Chapter 9

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 108 of 199

Fig.9.4 The state of the Working Memory after a FULL RESET

9.2 Storage of Measurement Results

While a monitoring task is being performed measurement data is

stored in the Monitor’s Display Memory. When the monitoring task is

complete the data in Display Memory is normally copied into the

Monitor’s Background Memory to prevent it from being overwritten

(and therefore lost) by measurement data stored during the next

monitoring task. The same amount of data can be stored in both the

Background Memory and the Display Memory.

Structure of the Background Memory:

The data from a maximum of 10 monitoring tasks can be stored in

the Background Memory provided the total amount of data from the

10 tasks does not exceed the total capacity of the Background

Memory. Each monitoring task stored in the Background Memory is

given a “location number” (from no. 1 to no. 10). The size of each

location is flexible and is equal to the space required to store all the

data collected during a monitoring task. Each monitoring task stored

will occupy at least 2 % of the total Memory capacity.