X-ON/X-OFF HANDSHAKE CONFLICT

The Monitor receives the

handshake signal from the

RS232 CTS ( clear to send)

line, but it has not been

set-up to use this CTS line

The Monitor and

RS232 device

(printer) do not

agree about which

handshake signal

they should use.

Perhaps the wrong

interface cable has

been used.

TIMEOUT. DEVICE NOT CONNECTED

Monitor wishes to send

data (data-log or

error-log) via the IEEE

interface, but there is no

response from the device

with the address entered

in the Monitor set-up

1 No device is con-

nected to the

Monitor

2 The device

address has

been entered in-

correctly in the

Monitor’s set-up.

TIMEOUT. DEVICE NOT READY

Monitor wishes to send

data

(print, data-log or error-

log) but after the start of

data transmission, the

device stops receiving

data.

1. The device is

“off-line“(it

cannot receive

data).

2. The device is not

able to receive

data fast

enough.

3. The device is no

longer

connected to

Monitor.