shown, press the INFO button. Text will appear on the screen to

indicate why this particular measurement has been marked.

The number which is shown on the display – between time and

concentration – is only “filled-in” after completion of a monitoring

task. The number indicates the order in which the measurements

were made. The “oldest” stored measurement is always given the

number 1 and the following measurements are numbered se-

quentially.

Overview of the Monitoring Task:

These displays provide an overview of the monitoring task set-up

parameters.

Start of the Monitoring Task:

The Display Memory has a finite size and therefore, if a monitoring

task runs for a long period of time, it will fill up all available memory

space, and data collected after the memory has been filled-up will

start to overwrite the data collected during the start of the

monitoring task. Some measurement results will therefore be lost

from the memory. In such a situation the date and time appearing

after MONITORING TASK STARTED….. will not be the same as the

first date and time displayed on the screen labelled

1