1-6

ADS FlowShark IS Manual

Only ADS approved sensors (ADS p/n 106186, 106194, and

3754-0001/0002/0003) shall be connected to the ADS Model

5000-IS1.

Due to the use of aluminum in the chassis of the ADS Model

5000-IS1, use special care when working with the product in

the hazardous area. Avoid any impact with the aluminum parts

(especially by rusted iron) that could produce a spark.

The only devices that may be connected to the COMM

Connector on the 5000-IS1 are the PC Communication Cable

(ADS p/n 106228) and the Communication Cable (ADS p/n

106227) connecting the monitor to the EMU Communication

Interface (ADS p/n 106226A). Only factory built

communication cables (ADS p/n 106227) shall be used.

In general, never rub any of the plastic parts (specifically, the

PVC connector potting boots and the velocity sensor housing)

with a dry cloth. This could result in the buildup of dangerous

static charges.