Maintenance and Troubleshooting 8-11

Refer to Replacing EMU Components on page 8-14 for instructions

on replacing the battery pack in the EMU (for low external voltage).

Checking the Sensors

Perform the following sensor inspections during regular site visits:

Verify that the installation ring or bands are secure and clear of

debris.

Clean the face of the ultrasonic sensor gently with a soft brush,

and wipe the sensor with a clean, moist cloth.

Verify that the face of the ultrasonic depth sensor is

horizontally level with the flow. If necessary, reposition the

sensor according to the procedures in Chapter 3, Sensor
Installation and Connection

.

Scrub the pressure sensor and face of the velocity sensor with a

soft bristle brush.

Confirm that all sensor cables are neatly arranged, securely

fastened, and free of debris, cuts, and breaks that may affect

performance. Replace any sensor that has a damaged cable.

Replacing the Pressure Depth Sensor Dryer Tub or
Desiccant

The pressure depth sensor’s dryer tube contains desiccant that

absorbs the moisture from the air traveling through the dryer tube

into the clear plastic tubing leading to the pressure depth sensor in

the pipe. Replace the dryer tube or the desiccant contained within

the dryer tube on a regular basis or as necessary. Desiccant can

absorb moisture only when it is blue in color. Pink desiccant

indicates that it can no longer absorb moisture. Therefore, if the

desiccant appears either colorless or pink upon inspection, replace

the dryer tube or desiccant immediately.