Maintenance and Troubleshooting 8-7

1. Contact a data analyst to collect the data remotely or collect the

data on-site. This will reduce the risk of data loss once the

battery is disconnected.

2. Remove the monitor from the hazardous area.
3. Remove the 4 bolts securing the monitor top and chassis to the

canister.

4. Wipe away any debris or moisture that could enter the chassis

once the inside of the unit is exposed.

5. Using the handle on the monitor top, pull the chassis straight up

from the canister. The battery pack sits at the bottom of the

canister, and has an 18-inch cable that allows the chassis to be

lifted just clear of the canister top.

Caution:

Be careful not to pull the chassis away from the

canister. This could break off the connector from the

voltage regulator PCB.

6. Place the chassis upright next to the canister opening,

disconnect the battery pack, and set the chassis aside.

Note:

Maintain the chassis in an upright position. The

chassis is top heavy and could fall over due to the weight

of the sensor cables and uneven ground surface.