Problem

Temperature reading from the pressure depth sensor
shows a value of -273

° C (-459° F).

Possible

Causes

Cable connecting the sensor to the monitor may be
loose or broken.

Sensor may be faulty.

Analog board is not responding or may be faulty.

Possible

Solutions

Secure the sensor connection to monitor (if necessary).

Replace the sensor.

Replace the monitor if defective.

Problem

Pressure depth readings are consistently incorrect.

Possible

Causes

Coefficients may be incorrect.

Pressure dryer tube may not be functioning properly.

Possible

Solutions

Retrieve or enter the correct coefficients, and reactivate
the monitor.

Inspect the color of the desiccant in the dryer tube, and
replace the tube (if necessary).

Problem

Pressure depth readings are consistently off by up to 3
inches (7.62 cm), but temperature readings are
accurate.

Possible

Cause

Monitor may be configured with an incorrect pressure
sensor offset.

Possible

Solution

Enter an offset to depth for consistency with
confirmation readings.