External power – ADS Environmental Services FlowShark QR 775003 A2 User Manual

Page 208

Advertising
background image

Maintenance and Troubleshooting 8-23

Problem

Data is missing at the beginning or end of the date
range following data collection.

Possible

Causes

Monitor activation may have failed.

Monitor time may be incorrect.

Monitor’s firmware file may be corrupt.

Possible

Solutions

Verify whether the monitor has been activated, and
activate it if necessary.

Check monitor time, and reset clock if necessary.

Run diagnostics in Profile to verify whether a firmware
problem may exist.

Contact your regional ADS representative.

Problem

An I/O error message displays when communicating
with the monitor.

Possible

Cause

Processor or analog board may be faulty.

Possible

Solutions

Re-attempt communication with the monitor.

Replace the monitor if defective.

Contact your regional ADS representative.

External Power

The following table contains troubleshooting techniques for

problems with the ADS External Power Supply. If another power

supply is in use, similar troubleshooting techniques may apply.

However, consult the manufacturer’s product documentation prior

to beginning the troubleshooting process.

Advertising