ADS Environmental Services FlowShark QR 775003 A2 User Manual

Page 207

Advertising
background image

8-22

ADS FlowShark IS Manual

Problem

Time on the monitor clock is incorrect.

Possible

Causes

Monitor clock may be a faulty.

PC clock may read incorrect time.

Possible

Solutions

Verify the time on the PC clock and correct if
necessary.

Reactivate the monitor to enable the clock.

Collect the data from the monitor and replace the
monitor if defective.

Problem

Profile provides a Device Time Out message.

Possible

Cause

Communications may be marginal.

Analog board may be faulty.

Possible

Solutions

Re-attempt communication with the monitor.

Replace the monitor if defective.

Contact your regional ADS representative.

Problem

Gap exists within the collected data.

Possible

Cause

Monitor time may be incorrect.

Monitor firmware file may be corrupt.

Possible

Solutions

Check monitor time, and reset clock if necessary.

Attempt to collect data within the gap.

Contact your regional ADS representative.