Sensor Installation and Connection 3-25

Warning:

Avoid over-tightening the ring. This could

bend the crank assembly.

7. Restore any silt moved to its previous level, and confirm that

the Doppler velocity sensor is still above the silt level.

8. Measure the physical offsets for the ultrasonic and pressure

depth sensors.
Ultrasonic Offset

Measure the distance from the crown

(top) of the pipe to the face of the ultrasonic depth sensor.

Pressure Offset

Measure the vertical distance from the

bottom of the pressure depth sensor to the bottom, center

of the pipe.

Note:

These values will be required when configuring the

monitor using the Profile software.

9. Secure the sensor cables from the ring to the future monitor

location in the manhole. Refer to Securing the Sensor Cables in
the Pipe and Manhole

on page 3-43 for more information.

Special Installations

A special installation requires two independent installations: one for

the ultrasonic depth sensor and one for the Doppler velocity and

pressure depth sensors.

Note:

Special installations do not involve spreader

mechanisms or rings. All hardware mounts directly to the

pipe surface with anchor bolts.