Getting Started Manual

922561-00 Rev A

119

3.

Touch the Fx tab to open the Function dialog, then touch Source1

and select a source trace from the pop-up.

4.

Touch the Phistogram tab at the right to open the Phistogram dialog.

5.

Touch Slice Direction and select Horizontal or Vertical slice from the

pop-up menu.

6.

Touch Slice Center and use the pop-up keypad to enter a value.

7.

Touch Slice Width and use the pop-up keypad to enter a value.

P

ERSISTENCE

T

RACE

R

ANGE

Using this math operator, you can enter the percent of the persistence trace
population to use in creating a new waveform.

P

ERSISTENCE

S

IGMA

Using this math operator, you can enter a scale, measured in standard
deviations, by which to create a new waveform.

Viewing Trends

This procedure explains how to view the Trend of a math function or
measurement parameter applied to a waveform. A Trend shows... See Track
vs. Trend
(on page 115) for a better understanding of the information a
Trend provides.

1.

In the menu bar, touch Measure → Measure Setup....

2.

Touch one of parameter tabs Px showing its corresponding dialog.

3.

Now, on the dialog, touch inside the Source1 data entry field and
select an input waveform from the pop-up menu.

4.

Touch inside the Measure data entry field and select a parameter
from the pop-up menu.

5.

Touch the Trend button at the bottom of the dialog.

Now, from the Math selection for Trend menu, select a math
function location Fx in which to store the Trend display. The Trend is

Advertising