Triggerscan – Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 98
 
LabMaster 10 Zi Oscilloscopes
90
922561-00 Rev A
TriggerScan
TriggerScan is a debugging tool (available for any trigger type) that helps 
you quickly find rare waveform glitches and anomalies. With TriggerScan, 
you can build a list of trigger setups to look for rare events and 
automatically sequence through each one. TriggerScan can use any type of 
trigger setup available including edge, width, and qualify as well as Smart 
Triggers (such as, glitch and runt triggers). TriggerScan automates two key 
processes in triggering rare events: 
Trains the system by looking at normal acquired waveforms. During 
the training, the oscilloscope analyzes the waveforms to determine 
what waveforms normally look like. Using this information, it 
generates a list of smart trigger setups to trigger on abnormal 
situations. 
Loads the smart trigger setups from the Trainer and cycles through 
these. As triggers occur, they are overlaid on the screen. All 
acquisition settings are preserved and you can use all the functions of 
the oscilloscope to find the root cause of these anomalies.