Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 135
 
Getting Started Manual
922561-00 Rev A
127
Running Pass/Fail Test
Initial Setup
1.
Touch Analysis → Pass/Fail Setup... on the menu bar.
2.
Touch the Actions tab.
3.
Touch the Enable Actions checkbox. This causes selected actions to 
occur on your waveform's passing or failing a test. 
4.
Touch the Summary View checkbox to enable a line of text (as 
follows) showing a concise status of the last waveform and keeping a 
running count of how many sweeps have passed. 
5.
Touch inside the Pass If control, and select a Boolean condition from 
the choices shown. 
6.
If you want to set up the test to end after a finite number of sweeps, 
touch the Stop Test checkbox. Then touch inside the After data entry 
control and provide a value using your preferred input control 
method. Additional information on using the touch screen controls 
can be found in the Dialog Area. 
7.
Use the Pass or Fail button on the If section of the dialog to set 
actions taking place when your waveform passes or fails the test. 
8.
The Then section of the dialog provides stop test, sound alarm, print 
result, emit pulse, or save (the waveform) actions. If you want to 
have the results printed (and your oscilloscope is not equipped with a 
printer) make sure your instrument is connected to a local or 
network printer. 
See Print Setup Dialog in the online Help.
9.
If you want to save your waveform automatically, touch the Save 
Setup... checkbox. This takes you out of the current dialog and opens 
the Save Waveform dialog. 
10.
Test Pass/Fail conditions by touching the Force Actions Once button. 
Press the Clear All button to quickly uncheck all checkboxes and 
make new selections.