Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 177
 
Getting Started Manual
922561-00 Rev A
169
Safety Compliance
EC
D
ECLARATION OF CONFORMITY
–L
OW
V
OLTAGE
The oscilloscope meets intent of EC Directive 2006/95/EC for Product 
Safety. Compliance was demonstrated to the following specifications as 
listed in the Official Journal of the European Communities: 
EN 61010-1:2010 Safety requirements for electrical equipment for 
measurement, control, and laboratory use – Part 1: General requirements 
EN 61010-2:030:2010 Safety requirements for electrical equipment for 
measurement, control, and laboratory use -- Part 2-030: Particular 
requirements for testing and measuring circuits. 
The design of the instrument has been verified to conform to the following 
limits put forth by these standards. 
Installation (Overvoltage) Categories:
Mains Supply Connector: CAT II, local distribution level, applicable to 
equipment connected to the mains supply (AC power source). 
Measuring Terminals: CAT 0, signal level, applicable to equipment 
measuring terminals connected to source circuits where measures 
are taken to limit transient voltages to an appropriately low level. 
Pollution Degree: 2, normally only dry, non-conductive pollution
occurs. Conductivity caused by temporary condensation should be 
expected. 
Protection Class: I, grounded equipment, protection against electric
shock is achieved by Basic Insulation and by means of a connection to 
the protective ground conductor in the building wiring. 
U.S.
N
ATIONALLY
R
ECOGNIZED
A
GENCY
C
ERTIFICATION
The oscilloscope has been certified by Underwriters Laboratories (UL) to 
conform to the following safety standard and bears UL Listing Mark 
1
:
UL 61010-1 Third Edition – Safety standard for electrical measuring and test 
equipment. 
1
UL certification does not extend to Channel Sync Mainframe Hub and