Vertical, Channel dialog – Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 75
 
Getting Started Manual
922561-00 Rev A
67
Vertical
You can access Vertical settings using the front panel Vertical controls, by 
selecting Vertical→ Channel Setup... on the menu bar, or by touching the 
Channel trace descriptor label. The following screen-shot shows the C1 
dialog on a LabMaster. For each open channel trace, a corresponding Cx 
dialog opens with setup controls. 
The Channel dialog also lets you perform certain actions on the channel 
trace, such as Math and Measure, without having to leave the dialog. 
NOTE
:
For LabMaster, the ChannelSYNC Calibration dialog provides another
vertical set up option.
Channel Dialog
The Cx tab of the dialog provides the following functions for the respective 
channel: 
Selected Input - For each channel, you can set the Channel Input to 
either Input A (Upper) or Input B (Lower). The Channel Row LED 
Indicators display which input is set for each channel. 
NOTE
: Not all oscilloscope models have Upper and Lower Channel Inputs. In
those cases, no Selected Input buttons are provided.
The Acquisition
Modules indicate alternative, or B channels, on named LED 
indicators.
Vertical Scale - For each channel, you can set the vertical scale or 
sensitivity and choose whether to use fixed or variable gain 
adjustment. 
Vertical Offset - For each channel, you can select between zero 
vertical offset or to set the offset to a specific value. 
Coupling - The maximum input voltage depends on the input used. 
Values are displayed on the front of the oscilloscope. Whenever the