Getting Started Manual

922561-00 Rev A

91

Training TriggerScan

The TriggerScan Trainer inspects a currently acquired waveform and
automatically builds a list of common trigger setups used to find rare
events.

NOTE

: Run the Trainer if you want to change the trigger types or if you change the

channel or signal. You must acquire and display at least 3 cycles of a signal before
running the Trainer.

1.

Touch Trigger → Trigger Setup... from the menu bar.

2.

On the Trigger dialog, click the TriggerScan tab.

3.

Touch the Trainer button and select a source channel from the pop-
up. Also choose the types of triggers the Trainer should use to train
the system and then touch the Start Training button. The training
begins. When it is complete, a list of smart trigger setups is displayed
in the Trigger List.