Getting Started Manual

922561-00 Rev A

77

1.

From the menu bar, choose Vertical > Channel <#> Setup. Touch the

ChannelSync Cal tab to the right of the C<#> tab.

2.

Place a checkmark next to the module(s) you want to calibrate and

touch the ChannelSync Cal button. The ChannelSync Calibration
Wizard opens.

3.

Select to measure Input A or Input B, or select Active Input to make
measurements on the active input of each channel.

4.

Select a Ref Channel to be used as a reference against which all other
measurements are made during your calibration.

5.

Enter the V/div (gain) setting for which you want to perform the
ChannelSync Cal on all channels. The default value is 100 mV/div,
which corresponds to the Fast Edge output amplitude.

6.

Connect the Fast Edge calibration signal (located on the front of the
Acquisition Module) to the reference input channel (in this example,
C1), then touch the Next button. The ChannelSync™ Cal begins.