Getting Started Manual

922561-00 Rev A

161

Absolute Volts - Have the vertical offset level indicator move with the
actual voltage level. With this selection, the waveform could move
off the grid.

NOTE

: Regardless of whether you select

Volts

or

Div

, the

Offset

shown in the

Channel Setup dialog always indicates volts. However, when

Div

is selected for the

Offset Control, the offset in volts is scaled proportional to the change in gain,
thereby keeping the division on the grid constant.

Delay Setting constant in - allows you to specify how Horizontal scale
changes are displayed. Choose either:

Time - As the timebase changes, keep the horizontal offset indicator
stationary. With this selection, the trigger point could move off the
grid.

Div(isions) - Have the horizontal offset indicator move with the
trigger point. With this selection, the trigger point remains on the
grid as you increase the timebase.

NOTE

: Regardless of whether you select Time or

Div

, the

Delay

shown in the

Timebase Setup dialog always indicates time.

Reset trigger counter before starting a new acquisition - If available, this
selection clears the trigger counter each time the oscilloscope issues an
ARM acquisition command and applies any trigger Holdoff conditions set on
the Trigger dialog: