Getting Started Manual

922561-00 Rev A

17

Now, connect the other end of the 1 Lane cable into the SYNC socket on the
back of the Acquisition Module.

Repeat these steps for every additional Acquisition Module in your system.

PLEASE NOTE THE FOLLOWING:

PCIe 1 Lane Cable plugs are keyed with a single groove along one
wide side
of the plug. The plug must be inserted into the socket with
the groove aligned properly.

Connect to your Acquisition Modules from the correct channel
groupings on the Master Control Module; meaning, your first
Acquisition Module is connected from the CHANNEL 1-4 output,
second from the CHANNEL 5-8 output, third from the CHANNEL 9-12
output, etc.

If you are connecting less than 5 Acquisition Modules, the PCIe 1
Lane Channel Sync Outputs
on the back of the Master Control
Module may not be skipped and must be connected in consecutive
order
into the PCIe 1 Lane Channel Sync Inputs on the back of the
corresponding acquisition modules.

Pull gently on the green pull tab on the cable plug for removal.