Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 86
 
LabMaster 10 Zi Oscilloscopes
78
922561-00 Rev A
As with System Skew calibration, you will see the Fast Edge output 
signal appear on the display and delay (skew) measurements will be 
made. 
7.
Connect the Fast Edge calibration signal to the next input (in this
example, C2), then touch the Next button.
Repeat Step 7 for each channel. As each channel is calibrated, its 
trace descriptor appears under the grid display area. 
8.
When you have finished all channels, a message indicates all
measurements have been performed. Touch the Finish button. The 
ChannelSync™ Calibration dialog shows details about all performed 
measurements and the deskew values applied. 
If you wish to perform additional ChannelSync calibrations on other 
gain settings, repeat the process from Step 5 with a different gain 
(V/div) setting and a Reference Correction value entered for that 
attenuator.