Signal Views
WaveScan provides distinct views of your signal:
Source View highlights all occurrences of edges that meet your 
criteria. 
Scan Overlay
places all captured edges one on top of the other in a
separate grid. You can apply monochromatic persistence in this view.
Scan Histogram
provides a statistical view of edges that meet your
search criteria.
Zoom View allows you to expand a waveform feature vertically and 
horizontally; to apply further processing (using Measure); to Store it, 
or to apply a descriptive Label to the feature. 
NOTE
: The number of grids varies from one to three depending on which views are
enabled. WaveScan handles this function automatically, and there is no option to 
move traces grid to grid, as would be the case under normal operation. 
Search Modes
Use the Mode control to select an appropriate Search to locate anomalies 
during acquisition. When each of the following modes are selected, a 
corresponding right-hand dialog provides additional controls. The following 
modes are available: 
Edge Mode – Designed for detecting the occurrence of edges; 
selectable slope and level. 
Non-monotonic Mode – For detecting threshold re-crosses; 
selectable slope, hysteresis, and level. 
Runt Mode – For detecting pulses failing to cross a threshold; 
selectable polarity and thresholds. 
Measurement Mode – For filtering and performing parameter 
measurements. 
Measurement Mode – Searches from 2 to 64 bits; ideal for bursted 
patterns where a PLL cannot lock. 
Bus Pattern Mode – Searches from 2 to 64 bits; enhances MSO