Trigger Front Panel Controls

Level – Sets the trigger level to 50%. Turn the knob to change the
trigger threshold level. The threshold level is indicated on the Trigger
label.

READY and TRIG'D Indicators - The READY indicator is lit when the
trigger is armed. TRIG'D is lit momentarily when a trigger occurs. A
fast trigger rate causes the light to stay lit continuously.

Setup - Opens the Trigger Setup... dialog. Corresponds with screen
menu selection: Trigger → Trigger Setup.... Press the Trigger S

ETUP

front panel button again to close the Trigger Setup... dialog.

Auto - Turn on Auto Trigger mode, which triggers the oscilloscope
after a time-out, even if the trigger conditions are not met.

Normal - Turn on Normal Trigger mode, which triggers the
oscilloscope each time a signal is present that meets the conditions
set for the type of trigger selected.

Single - Turn on Single Trigger mode, which arms the oscilloscope to
trigger once (single-shot acquisition) when the input signal meets the
trigger conditions set for the type of trigger selected. If the scope is
already armed, it will force a trigger.

Stop Prevent the scope from triggering on a signal. If you boot up the
instrument with the trigger in Stop mode, a no trace available
message is shown. Press the Trigger A

UTO

front panel button to

display your trace.