LabMaster 10 Zi Oscilloscopes
88
922561-00 Rev A
Edge Trigger with Holdoff by Time. The broken upward-pointing arrows indicate 
potential triggers, which would occur if other conditions are met. The bold arrows 
indicate where the triggers actually occur when the holdoff time has been exceeded. 
H
OLD
O
FF BY
E
VENTS
Select a positive or negative slope and a number of events. An event is the 
number of times the trigger condition is met after the last trigger. A trigger 
is generated when the condition is met after this number, counted from the 
last trigger. The count is restarted on each trigger. For example, if the event 
number is two, the trigger occurs on the third event. 
Edge Trigger with Holdoff by Events (in this example, two events). The bold edges on 
the trigger source indicate that a positive slope has been selected. The broken, 
upward-pointing arrows indicate potential triggers, while the bold ones show where 
triggers actually occur after the holdoff expires.