LabMaster 10 Zi Oscilloscopes

88

922561-00 Rev A

Edge Trigger with Holdoff by Time. The broken upward-pointing arrows indicate
potential triggers, which would occur if other conditions are met. The bold arrows
indicate where the triggers actually occur when the holdoff time has been exceeded.

H

OLD

O

FF BY

E

VENTS

Select a positive or negative slope and a number of events. An event is the
number of times the trigger condition is met after the last trigger. A trigger
is generated when the condition is met after this number, counted from the
last trigger. The count is restarted on each trigger. For example, if the event
number is two, the trigger occurs on the third event.

Edge Trigger with Holdoff by Events (in this example, two events). The bold edges on
the trigger source indicate that a positive slope has been selected. The broken,
upward-pointing arrows indicate potential triggers, while the bold ones show where
triggers actually occur after the holdoff expires.