The Amplitude and Range of the trigger level have limits as detailed
in the datasheet specifications regularly maintained at
www.teledynelecroy.com.

Touch inside the Level data entry control and provide a value using
your preferred input control method. Additional information on using
the touch screen controls can be found in the Dialog Area.

Provide a voltage level (in millivolts).

NOTE

: Once specified,

Trigger Level

and

Coupling

are the only parameters

remaining unchanged as you switch from trigger mode to trigger mode for each
trigger source.

Optimize for HF

The Optimize for HF checkbox can be used on an Edge trigger to reject high
or low frequencies.

The checkbox is marked by default; meaning, the instrument is optimized
for high-frequency waveforms.

NOTE

: If you are measuring a waveform that is 10 MHz or slower, be sure to

unmark the checkbox to avoid triggering on an incorrect slope.

Width Condition

Width triggers (and other triggers equipped with conditional logic) have
conditional logic settings as follows.

Your Width Condition can be specified as Less Than, Greater Than, In Range
and Out Of Range.