The Amplitude and Range of the trigger level have limits as detailed 
in the datasheet specifications regularly maintained at 
www.teledynelecroy.com. 
Touch inside the Level data entry control and provide a value using 
your preferred input control method. Additional information on using 
the touch screen controls can be found in the Dialog Area. 
Provide a voltage level (in millivolts).
NOTE
: Once specified,
Trigger Level
and
Coupling
are the only parameters
remaining unchanged as you switch from trigger mode to trigger mode for each 
trigger source. 
Optimize for HF
The Optimize for HF checkbox can be used on an Edge trigger to reject high 
or low frequencies. 
The checkbox is marked by default; meaning, the instrument is optimized 
for high-frequency waveforms. 
NOTE
: If you are measuring a waveform that is 10 MHz or slower, be sure to
unmark the checkbox to avoid triggering on an incorrect slope.
Width Condition
Width triggers (and other triggers equipped with conditional logic) have 
conditional logic settings as follows. 
Your Width Condition can be specified as Less Than, Greater Than, In Range 
and Out Of Range.