Channelsync mainframe hub – Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 31
 
Getting Started Manual
922561-00 Rev A
23
ChannelSync Mainframe Hub
The CMH-20 ChannelSync Mainframe Hub provides a simple and effective 
means to expand a LabMaster system beyond 20 channels (5 Acquisition 
Modules). This is accomplished without any degradation of the timing 
accuracy specifications. The Mainframe Hub is populated with cards for 
each acquisition module (the back panel of the CMH-20 shown below is 
populated with 20 cards for use with 20 Acquisition Modules, or 80 total 
channels). 
Connect a PCIe 1 Lane cable from the MCM-Zi Channel 1-4 output to the 
single PCIe 1 Lane input at the lower right of the CMH-20 back panel. 
Connect four PCle 4 Lane cables from the MCM-Zi data outputs for 
channels 1-4, 5-8, 9-12, and 13-16 to the four PCIe 4 Lane data inputs along 
the bottom of the CMH-20 back panel (to the left of the PCLe 1 input). 
Connect from the PCIe 1 Lane Control, PCIe 4 Lane Data, and 10 GHz 
ChannelSync Clock outputs to the corresponding inputs on the Acquisition 
Modules.