Power cable and main power switch – Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual

Page 32

Advertising
background image

LabMaster 10 Zi Oscilloscopes

24

922561-00 Rev A

Power Cable and Main Power Switch

Teledyne LeCroy advises that you note the power ratings of the individual
modules and connect them to suitably rated circuits.

MCM-Zi Control Module comes with an IEC60320 Type C13 power
inlet connector.

LM10xx-Zi Acquisition Module comes with an IEC60320 Type C20
power inlet connector.

CAUTION: The combined draw from your Master Control Module
and Acquisition Modules (max of 20) can be many kilowatts. Two

acquisition modules connected to a single circuit may cause an overcurrent
condition for that circuit.

When all connections are made, the Standby Power switch on the front of
the Master Control Module powers on all components as a single unit.

NOTE

: If any connections are incorrect, the switch will not power on the system.

Always refer to the product datasheet at www.teledynelecroy.com for the
most current and detailed specifications.