Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual

Page 149

Advertising
background image

Getting Started Manual

922561-00 Rev A

141

Recalling Oscilloscope Setup(s)

Access the Recall Setup... dialog by either selecting File → Recall Setup...
from the menu bar or clicking the Recall Instrument Setup button on the
main Save/Recall dialog. Use one of the following two methods to save
setup(s).

R

ECALL

I

NTERNAL

S

ETUP

(

S

)

FROM

Y

OUR

I

NSTRUMENT

Touch one of the six Recall buttons in the Recall From Internal Setup...
section of the dialog.

Each Recall button is labeled with a corresponding setup storage slot. If a
setup is stored, SetupX is shown along with the date/time of the save,
otherwise, the slot is labeled Empty.

R

ECALL

S

ETUP

(

S

)

D

IRECTLY FROM A

F

ILE

Recall your setup(s) directly from a file by using controls on the Recall From
File
section on the right side of the dialog. Touch inside the Recall panel
from file
data entry control and provide a path to the destination folder. Or
touch Browse to navigate to the destination folder. Finally, touch the Recall
Now
button.

Restoring Default Settings

To reset the settings to the default configuration, press the front panel
Default Setup button or the Recall Setup dialog Recall Default button.

Following are some of the default settings for your oscilloscope:

Channels 1 and 2 are turned on at 50 mV/div, 0V offset, linear
interpolation

All other measurements are cleared for Math, Reference Waveforms,
and Channels

Trigger is set to Channel 1 with an Auto positive edge, DC coupled,
and with a 0V level

The Timebase set to 50 ns/div, 10ks, 20 GS/s, and with zero delay

Smart memory is set to Set max memory with 100ks selected

The Cursors are turned off

The Grids are set to AutoGrid

Advertising