LabMaster 10 Zi Oscilloscopes

148

922561-00 Rev A

address if your network supports DHCP. If it does not, you can assign
a static address in the standard Windows network setup menu.

The Remote Control Assistant monitors communication between your PC
and oscilloscope when you are operating the instrument remotely. You can
log all events or errors only. This log can be invaluable when you are
creating and debugging remote control applications.

Remote Control Setup

If you are connecting the oscilloscope to a network, first contact your
Information Systems administrator. If you are connecting the oscilloscope
directly to your PC, connect a GPIB or Ethernet cable between them.

1.

On the Remote dialog, make a Port selection: TCPIP (transmission
control protocol/Internet protocol) or GPIB (general purpose
interface bus). If you do not have a GPIB card installed, the GPIB
selection will not be accessible.

2.

If you are using GPIB, set a GPIB address by touching inside the GPIB
Address
data entry control and provide an address.
OR

If you are using TCPIP, press the Net Connections button; the
Windows Network Connections window opens. You can reconfigure
the oscilloscope's connection if it is already connected to the network
or make a new connection.

NOTE

: Your instrument allows you to restrict remote control access to certain

clients. To restrict access, under

Security

, touch the

Yes

button and enter the IP

addresses or domain name server names you want to restrict separated by a
comma.