Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual

Page 49

Advertising
background image

Getting Started Manual

922561-00 Rev A

41

Channels – Control channel ON/OFF and which channel is active for

the Vertical Offset and Volts/Div knobs controls. If a channel is OFF,
pressing that channel button turns it on and makes it active. If a
channel button is ON, pressing that channel button makes it active,
and then pressing it a second time turns it OFF.

Offset - Toggles between a zero vertical offset value and the previous
vertical offset value for the selected channel. Turn to change the
vertical offset value for the selected channel.

Gain - Toggles between fixed and variable gain adjustment. Turn to
change the gain value.

NOTE:

Front panel Channel buttons are only lit when Channels 1 through 4 are

active; however, Gain and Offset knobs will control any channels from any
connected acquisition module. Touch the display Trace Descriptor Label to activate
the channel before using the front panel knobs to make desired Gain and Offset
adjustments.

Zoom and Math Front Panel Controls

NOTE: Zoom and Math front panel controls correspond with screen menu selection:
Math → Zoom Setup....

Horizontal Position - Resets the horizontal zoom position to zero.
Turn to change the horizontal position of the selected math or zoom
trace.

Horizontal Ratio - Toggles between fixed and variable horizontal
zoom ratio adjustment. Turn to change the horizontal zoom ratio of
the selected math trace.

Advertising