Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 39
 
Getting Started Manual
922561-00 Rev A
31
Most active probes match probe to oscilloscope response
automatically using probe response data stored in an on-board 
EEPROM. This ensures the best possible combined probe plus 
oscilloscope channel frequency response without the need to 
perform any de-embedding procedure. 
Interfaces differ in bandwidth capabilities, so the interfaces contained on 
your oscilloscope—and their exact location—depend on the bandwidth 
rating of the oscilloscope model you purchased. 
High-bandwidth interfaces ( 45 GHz and up) must be enabled by selecting 
the appropriate Dual Bandwidth Interleave setting on the Timebase dialog 
and the Input B setting on the Channel dialog. 
ProBus Interface
LabMaster 10xxZi Acquisition Modules rated to 36 GHz use the ProBus 
interface for AUX In and AUX Out connections. These connectors are 
located on the front of the unit next to the channel inputs. 
The ProBus interface contains a 6-pin power and communication connection 
and a BNC signal connection to the probe. It offers both 50 Ω/1 MΩ input 
impedance and provides probe power and control for a wide range of 
probes such as high impedance passive probes, high impedance active 
probes, current probes, high voltage probes, and differential probes. ProBus 
also includes sense rings for detecting passive probes. The ProBus interface 
may also have a BNC-terminated cable connected directly to it. 
ProBus is based on a BNC connector and, depending on the exact BNC 
connector used and the oscilloscope design, is rated for up to 4 GHz with 50 
Ω coupling or up to 1 GHz for 1 MΩ coupling. 
2.92 mm Interface
All LabMaster 10xxZi Acquisition Modules use the 2.92 mm interface to 
input signals up to 36 GHz on channels 1-4, input A. This interface consists 
of a precision connector and a LEMO power and communication connector. 
It offers 50 Ω input impedance only.