Signal Inputs
Probes
Teledyne LeCroy offers a variety probes for use with your oscilloscope. The 
types compatible with the LabMaster 10 Zi are listed below. Visit 
teledynelecroy.com for specifications and ordering information. 
Single-Ended - A single-ended, active probe is associated with 
measuring voltages at high frequencies. Measurement with an active 
probe requires a test point and a ground point. The ground (also 
called earth) acts as a zero reference for the test point measurement. 
These probes are typically low-bandwidth and require an adapter to 
connect to the 2.92 mm and 1.85 mm interfaces on the 10-xxZi 
Acquisition Modules. 
Differential Probes - Differential active probes are like two probes in 
one. Instead of measuring a test point in relation to a ground point 
(like single-ended active probes), differential probes measure the 
difference in voltage of a test point in relation to another test point. 
Some models of differential probes may require an adapter to 
connect to the 2.92 mm and 1.85 mm interfaces on the 10-xxZi 
Acquisition Modules. 
Interfaces
LabMaster 10 Zi systems are equipped with a variety of interfaces to allow 
you to connect cables directly to the oscilloscope channels, reducing the 
need to use costly adapters that may be easily lost or misplaced. 
The inputs power the probe and completely integrate the probe with the 
oscilloscope channel in a number of ways: 
Upon connection, the probe is recognized and some setup
information, such as input coupling and attenuation, is performed 
automatically. 
System (probe plus oscilloscope) gain settings are automatically
calculated and displayed based on the probe attenuation.