Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 84
 
LabMaster 10 Zi Oscilloscopes
76
922561-00 Rev A
If you perform ChannelSync Cal on multiple gain settings, perform it 
first on the 100 mV/div setting, then perform it on other gain 
settings (with “Reference Correction” values entered as appropriate 
for those other gain settings.) 
For the highest possible accuracy across the full input range of the 
oscilloscope, a ChannelSync Cal should be performed on at least one 
gain setting in each of the following gain ranges: 
2.92mm up-to-36 GHz Inputs
1.85mm up-to-65 GHz Inputs
≤10 mV/div
≤10 mV/div
10.1 to 11.8 mV/div
10.1 to 11.8 mV/div
11.9 to 17.0 mV/div
11.9 to 14.2 mV/div
17.1 to 20.5 mV/div
14.3 to 17.0 mV/div
20.6 to 30.0 mV/div
17.1 to 20.5 mV/div
30.1 to 39.5 mV/div
20.6 to 25.0 mV/div
39.6 to 50.0 mV/div
25.1 to 30.0 mV/div
50.1 to 75.0 mV/div
30.1 to 39.5 mV/div
75.1 to 90.0 mV/div
39.6 to 50.0 mV/div
90.1 to 100.0 mV/div
50.1 to 63.0 mV/div
100.1 to 126.0 mV/div
63.1 to 80 mV/div
126.1 to 158.0 mV/div
158.1 to 205.0 mV/div
205.1 to 250.0 mV/div
250.1 to 315.0 mV/div
315.1 to 500 mV/div
If you add cables or probes to your setup subsequent to running the 
ChannelSync Cal, you must manually deskew these items in the 
Channel dialog to account for their propagation delays. The values 
you provide on the Channel dialog (in the deskew section) are 
additional to values shown on the ChannelSync Calibration dialog.