LabMaster 10 Zi Oscilloscopes

76

922561-00 Rev A

If you perform ChannelSync Cal on multiple gain settings, perform it
first on the 100 mV/div setting, then perform it on other gain
settings (with “Reference Correction” values entered as appropriate
for those other gain settings.)

For the highest possible accuracy across the full input range of the
oscilloscope, a ChannelSync Cal should be performed on at least one
gain setting in each of the following gain ranges:

2.92mm up-to-36 GHz Inputs

1.85mm up-to-65 GHz Inputs

≤10 mV/div

≤10 mV/div

10.1 to 11.8 mV/div

10.1 to 11.8 mV/div

11.9 to 17.0 mV/div

11.9 to 14.2 mV/div

17.1 to 20.5 mV/div

14.3 to 17.0 mV/div

20.6 to 30.0 mV/div

17.1 to 20.5 mV/div

30.1 to 39.5 mV/div

20.6 to 25.0 mV/div

39.6 to 50.0 mV/div

25.1 to 30.0 mV/div

50.1 to 75.0 mV/div

30.1 to 39.5 mV/div

75.1 to 90.0 mV/div

39.6 to 50.0 mV/div

90.1 to 100.0 mV/div

50.1 to 63.0 mV/div

100.1 to 126.0 mV/div

63.1 to 80 mV/div

126.1 to 158.0 mV/div

158.1 to 205.0 mV/div

205.1 to 250.0 mV/div

250.1 to 315.0 mV/div

315.1 to 500 mV/div

If you add cables or probes to your setup subsequent to running the
ChannelSync Cal, you must manually deskew these items in the
Channel dialog to account for their propagation delays. The values
you provide on the Channel dialog (in the deskew section) are
additional to values shown on the ChannelSync Calibration dialog.