Using the spo, Parameter control panel, Setting upper and lower systolic alarm limits – ZOLL Propaq M Rev B User Manual

Page 104

Advertising
background image

C

HAPTER

9 M

ONITORING

P

ULSE

O

XIMETRY

(S

P

O2)

9-8

www.zoll.com

9650-0803-01 Rev. B

Setting Upper and Lower Systolic Alarm Limits

Initially, the SpO2 Alarm Settings menu specifies whether the SpO2 alarms are enabled (

ON

)

or

disabled (

OFF

),

and displays the default upper and lower systolic limits. The upper and lower

limits can be

ON

or

OFF

(default is

OFF

).The following table lists the default SpO

2

limits for

adult, pediatric, and neonate patients, and gives the range in which you can set these limits:

Using the SpO

2

Parameter Control Panel

Use the navigation keys to highlight and select the SpO

2

numeric display to display the

SpO

2

Parameter Control Panel, on which you can select options that optimize the use of

SpO

2

for the patient:

Figure 9-4 SPO

2

Parameter Control Menu

Patient Type

SpO

2

Limit Default

SpO

2

Limit Range

Adult

Lower: 85%

Upper: 100%

Lower: 50 - 98%

Upper: 52 - 100%

Pediatric

Lower: 85%

Upper: 100%

Lower: 50 - 98%

Upper: 52 - 100%

Neonate

Lower: 85%

Upper: 95%

Lower: 50 - 98%

Upper: 52 - 100%

0 to 30 mmHg

0 to 150 mmHg

CO2 Scale

0 to 100 mmHg

0 to 60 mmHg

SpO2

Lower

Upper

SpO2 Alarm

85

100

Response

Normal

HR/PR Tone

Low