ZOLL Propaq M Rev B User Manual

Page 94

Advertising
background image

C

HAPTER

8 M

ONITORING

N

ON

-I

NVASIVE

B

LOOD

P

RESSURE

(NIBP)

8-12

www.zoll.com

9650-0803-01 Rev. B

On the NIBP Parameter Control Panel, you can select an alarm configuration option (

SYS

,

DIA

, or

MAP

) to enable/disable alarms and set high and low alarm limits (as described

previously).

You can also configure these NIBP options:

NIBP Mode (

Mode

)

NIBP Automatic Measurement Interval (

Auto Interval

)

Enable/Disable Smartcuf motion artifact filter

(Smartcuf)

Start/Stop Turbocuf Measurements (

Turbocuf

)

Specify NIBP Mode

You can specify that the Propaq M unit operate in either Manual or Automatic Mode.

In Manual Mode, the Propaq M unit takes a single NIBP measurement when you press the

NIBP button on the front panel (

). To repeat the NIBP measurement, you must press the

NIBP key again (select

Manual

).

In Automatic Mode, the Propaq M unit takes the first of a series of NIBP measurements when
the

Auto Interval

timer expires, and then repeats the NIBP measurement at this specified

interval.

Note:

You can press the NIBP button and manually initiate an NIBP reading at any time
while in Automatic Mode. Manually initiated NIBP readings will not affect the timing
of subsequent NIBP readings in Automatic Mode.

Specify the Automatic NIBP Measurement Interval

You can specify the time interval between NIBP measurements in Automatic Mode. The
default interval between measurements is 5 minutes.

You can specify intervals of

1, 2, 3, 5, 10, 15, 30,

and

60

minutes between NIBP measurements.

Enable/Disable Smartcuf Motion Artifact Filter

Enabling the use of the Smartcuf motion artifact filter increases the accuracy of NIBP
measurements when motion artifact or weak pulses are present.

NIBP measurements can be adversely affected by many factors, such as cardiac arrthythmias,
sudden changes in blood pressure, body motions such as shivering and convulsions, bumping
the cuff, vibration, vehicle motion, or weak pulses. The Smartcuf filter synchronizes the NIBP
reading with the R-wave of the patient’s ECG to eliminate noise created by patient motion or
vibration.