Recommended Minimum Preventive Maintenance Schedule

9650-0803-01 Rev. B

Propaq M Operator’s Guide

15-3

Cleaning Cables and Accessories

Cables, cuff tubing, and other accessories can be wiped clean with a damp cloth moistened in a
mild detergent solution or according to manufacturer's instructions.

Recommended Minimum Preventive Maintenance
Schedule

Operational Tests should be performed at regular intervals. The Operational Tests augment the
automated self-tests that the Propaq M unit performs to help ensure readiness. The following
Operational Tests may be performed by the user:

Keypad Test

LED Test

LCD Test

RFU Indicator Test

Audio Test

Additional Tests may be performed by a qualified Biomedical Technician (BMET):

Battery Test

Fan Test

NIBP Test

USB Test

The Operational Test Procedures are described in the Propaq M Service Manual.

In addition to inspection and cleaning instructions, the following is recommended:

Perform NIBP calibration check

Perform CO

2

calibration check

Perform Functional Verification Tests

These tests shall be performed every year or according to local requirements (to be performed
by a qualified Biomedical Equipment Technician (BMET)) as described in the Propaq M
Service Manual
.