Enabling/Disabling IBP Alarms and Setting Alarm Limits

9650-0803-01 Rev. B

Propaq M Operator’s Guide

11-7

Setting Upper and Lower Diastolic (DIA) Alarm Limits

Initially, the IBP Diastolic Alarm Settings menu specifies that the IBP diastolic pressure alarms
are enabled (

ON

)

or disabled (

OFF

),

and displays the default Upper and Lower diastolic limits.

The upper and lower limits can be

ON

or

OFF

(default is

OFF

). The following table lists the

default diastolic limits for adult, pediatric, and neonate patients, and gives the range in which
you can set these limits:

Setting Upper and Lower Mean Arterial Pressure (MEAN) Alarm Limits

Initially, the IBP MEAN Alarm Settings menu specifies that IBP MEAN alarms are enabled
(

ON

)

or disabled (

OFF

),

and displays the default Upper and Lower MEAN limits. The upper

and lower limits can be

ON

or

OFF

(default is

OFF

). The following table lists the default MEAN

alarm limits for adult, pediatric, and neonate patients, and gives the range in which you can set
these limits:

Patient Type

IBP Diastolic Limit Default

IBP Diastolic Limit Range

Adult

Lower: 35 mmHg

Upper: 110 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg

Pediatric

Lower: 35 mmHg

Upper: 100 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg

Neonate

Lower: 30 mmHg

Upper: 70 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg

Patient Type

IBP Mean Limit Default

IBP Mean Limit Range

Adult

Lower: 50 mmHg

Upper: 120 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg

Pediatric

Lower: 50 mmHg

Upper: 110 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg

Neonate

Lower: 35 mmHg

Upper: 80 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg