Lead display options, Specifying the 12-lead frequency response – ZOLL Propaq M Rev B User Manual
Page 143
12-Lead Display Options
9650-0803-01 Rev. B
Propaq M Operator’s Guide
13-9
12-Lead Display Options
The Propaq M unit provides additional 12-Lead display options that you can specify through
the Supervisor parameter control panels (access to Supervisor is passcode-controlled).
Press the Setup quick access key (
), then select Supervisor
.
Using the navigation keys,
highlight and select the four digits in the Supervisor passcode:
Select the Supervisor> ECG> 12-Lead options to display the 12-Lead parameter control panel:
Specifying the 12-Lead Frequency Response
This option allows you to specify the frequency response of the 12-Lead waveform display.
You can specify the following trace display ranges:
12-Ld Frequency Response
Diagnostic
Display Type
Frequency Response
Diagnostic
0.05 to 150 Hz
Filtered Diagnostic
0.25 to 40 Hz