background image

12-Lead Display Options

9650-0803-01 Rev. B

Propaq M Operator’s Guide

13-9

12-Lead Display Options

The Propaq M unit provides additional 12-Lead display options that you can specify through
the Supervisor parameter control panels (access to Supervisor is passcode-controlled).

Press the Setup quick access key (

), then select Supervisor

.

Using the navigation keys,

highlight and select the four digits in the Supervisor passcode:

Select the Supervisor> ECG> 12-Lead options to display the 12-Lead parameter control panel:

Specifying the 12-Lead Frequency Response

This option allows you to specify the frequency response of the 12-Lead waveform display.

You can specify the following trace display ranges:

12-Ld Frequency Response

Diagnostic

Display Type

Frequency Response

Diagnostic

0.05 to 150 Hz

Filtered Diagnostic

0.25 to 40 Hz