Temperature
9650-0803-01 Rev. B
Propaq M Operator’s Guide
A-9
Transducer:
Sensitivity: 5uV/V/mmHg
Offset: +/- 125 mmHg including transducer offset
Excitation Impedance Range: 150 to 10,000 ohms
Excitation Voltage: 4.75 +/- 0.25 VDC
Connector: 6-pin circular MS3100 series
Temperature
Number of Channels: 2
Measurement Range: 0° to 50° C
Accuracy:
± 0.1° C from 10° C to 50° C
± 0.2° C from 0° C to 10° C
Resolution: 0.1° C
Scale: Fahrenheit or Celsius.
Temperature Display Signal: 20Hz, no averaging.
Probe: YSI 400 and 700 series
Display: T1, T2, T
Minimum Measurement Time: See the probe’s Instructions for Use to obtain minimum
measurement times for accurate readings. The Propaq M does not add any clinically significant
time to obtain accurate readings.
Connect to:
A
B
C
D
E
Signal Type
Sig (-)
Exc (+)
Sig (+)
Exc (-)
shield