Temperature

9650-0803-01 Rev. B

Propaq M Operator’s Guide

A-9

Transducer:

Sensitivity: 5uV/V/mmHg

Offset: +/- 125 mmHg including transducer offset

Excitation Impedance Range: 150 to 10,000 ohms

Excitation Voltage: 4.75 +/- 0.25 VDC

Connector: 6-pin circular MS3100 series

Temperature

Number of Channels: 2

Measurement Range: 0° to 50° C

Accuracy:

± 0.1° C from 10° C to 50° C

± 0.2° C from 0° C to 10° C

Resolution: 0.1° C

Scale: Fahrenheit or Celsius.

Temperature Display Signal: 20Hz, no averaging.

Probe: YSI 400 and 700 series

Display: T1, T2, T

Minimum Measurement Time: See the probe’s Instructions for Use to obtain minimum
measurement times for accurate readings. The Propaq M does not add any clinically significant
time to obtain accurate readings.

Connect to:

A

B

C

D

E

Signal Type

Sig (-)

Exc (+)

Sig (+)

Exc (-)

shield