Chapter 4 trends, Displaying the trends status window – ZOLL Propaq M Rev B User Manual
Page 51
9650-0803-01 Rev. B
Propaq M Operator’s Guide
4-1
Chapter 4
Trends
The Propaq M unit accumulates a patient’s trend information by logging all monitored vital
sign measurements to memory at a user-configurable interval. It also logs all monitored vital
sign measurements whenever
•
an NIBP measurement is taken and the Trend on NIBP option is on
•
when you press the Snapshot button (
)on the front panel
•
when a patient alarm occurs and the Trend on Alarm option is on
The Propaq M unit can store at least 24 hours of trend information when logged at a 1 minute
trend interval. You can view or save to external memory all logged trend information.
Displaying the Trends Status Window
The Propaq M unit displays the logged trend information in the Trends status window. Press the
Display/Home button (
) to display the Trends window, the primary ECG trace, and the
small numeric displays for each monitoring function: