Batteries – ZOLL Propaq M Rev B User Manual

Page 16

Advertising
background image

C

HAPTER

1 G

ENERAL

I

NFORMATION

1-8

www.zoll.com

9650-0803-01 Rev. B

Batteries

Propaq M models use an easily replaced rechargeable lithium-ion battery pack (the Propaq
Battery Pack). A new, fully charged battery pack typically delivers more than 8 hours of ECG
monitoring. Use of other options (such as higher screen brightness or shorter NIBP intervals)
reduces this time.

When a LOW BATTERY icon appears on the display and the unit emits three beeps in
conjunction with the displayed battery icon, the battery must be replaced and recharged.

You can charge the battery by either of the following methods:

Internal charging — plug the Propaq M into an auxiliary power adapter to automatically
begin charging the installed battery pack. The front panel battery indicator operates as
follows:

Note:

Upon power up, it takes approximately 45 seconds for the LEDs on the battery to
accurately display run time.

External charging — use the ZOLL SurePower™ Battery Charger with the Propaq M
battery adapter to charge the battery pack and test the battery’s capacity. For details, refer to
the Propaq Battery Pack Guide.

The Recalibration LED icon (

) lights for approximately 10 seconds (after you press and

release the Display button) if the battery needs to be calibrated. If the Recalibration LED lights,
the runtime indicator will not display run time for that battery. For best performance of the
battery, you should recalibrate the battery as soon as possible.

To manually recalibrate the SurePower Battery Pack, you can insert the battery into the
SurePower Charger Station and perform a Manual Test (for more information, see the ZOLL
SurePower Charger Station Operator’s Guide).

After you recalibrate the battery, the Recalibration LED will only flash when you press the
Display button.

When the indicator is:

It means:

Steady yellow

Battery is charging.

Steady green

Battery is charged.

Alternating yellow and
green

The charge state cannot be
determined or a battery charging
fault has been detected.

Not lit

No battery in device.