Guidance and Manufacturer’s Declaration – Electromagnetic Immunity (IEC
60601-1-2 Table 203)

The Propaq M unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the Propaq M should assure that it is used in such an environment.

Immunity Test

IEC 60601 Test Level

Compliance
Level

Electromagnetic Environment – Guidance
IEC 60601-1-2

Portable and mobile RF communications
equipment should be used no closer to any
part of the Propaq M unit, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.

Recommended Separation Distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 MHz
outside ISM bands

a

3 Vrms

d = 1.17

3 Vrms
150 kHz to 80 MHz in
ISM bands

a

3 Vrms

d = 1.17

Radiated RF
IEC 61000-4-3

10 V/m
80 MHz to 2.5 GHz

10 V/m

d = 1.20
80 MHz to 800 MHz

d = 2.30
800 MHz to 2.5 GHz

3 V/m
80 MHz to 2.5 GHz
(IEC 60601-2-34)

3 V/m
(IBP only)

d = 4
80 MHz to 800 MHz

d = 7.67
800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m).

b

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey

c

, should be

less than the compliance level in each frequency range

d

.

Interference may occur in the vicinity of equipment marked with the following symbol: