ZOLL Propaq M Rev B User Manual

Page 92

Advertising
background image

C

HAPTER

8 M

ONITORING

N

ON

-I

NVASIVE

B

LOOD

P

RESSURE

(NIBP)

8-10

www.zoll.com

9650-0803-01 Rev. B

6.

When you are finished changing values on the alarm settings menu, select the back arrow to
exit the menu.

Setting Upper and Lower Systolic Alarm Limits

Initially, the NIBP Systolic Alarm Settings menu specifies that the NIBP systolic pressure
alarms are enabled (

ON

)

or disabled (

OFF

),

and displays the default upper and lower systolic

limits. The upper and lower limits can be

ON

or

OFF

(default is

OFF

). The following table lists

the default NIBP Systolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:

Setting Upper and Lower Diastolic Alarm Limits

Initially, the NIBP Diastolic Alarm Settings menu specifies that the NIPBP diastolic pressure
alarms are enabled (

ON

)

or disabled (

OFF

),

and displays the default Upper and Lower diastolic

limits. The upper and lower limits can be

ON

or

OFF

(default is

OFF

). The following table lists

the default diastolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:

Patient Type

NIBP Systolic Limit Default

NIBP Systolic Limit Range

Adult

Lower: 75 mmHg

Upper: 220 mmHg

Lower: 30-258 mmHg

Upper: 32-260 mmHg

Pediatric

Lower: 75 mmHg

Upper: 145 mmHg

Lower: 30-158 mmHg

Upper: 32-160 mmHg

Neonate

Lower: 50 mmHg

Upper: 100 mmHg

Lower: 20-118 mmHg

Upper: 22-120 mmHg

Patient Type

Diastolic Limit Default

Diastolic Limit Range

Adult

Lower: 35 mmHg

Upper: 110 mmHg

Lower: 20-218 mmHg

Upper: 22-220 mmHg

Pediatric

Lower: 35 mmHg

Upper: 100 mmHg

Lower: 20-128 mmHg

Upper: 22-130 mmHg

Neonate

Lower: 30 mmHg

Upper: 70 mmHg

Lower: 10-108 mmHg

Upper: 12-110 mmHg