Setting CO2 and Respiration Rate Alarms

9650-0803-01 Rev. B

Propaq M Operator’s Guide

10-9

Setting Upper and Lower EtCO

2

Limits

Initially, the EtCO2 Alarm menu specifies that the EtCO

2

alarms are enabled (

ON

) or disabled

(

OFF

), and displays the default Upper and Lower EtCO

2

Limits. The upper and lower limits can

be

ON

or

OFF

(default is

OFF

). The following table lists the default EtCO

2

limits for adult,

pediatric, and neonate patients, and gives the range in which you can set these limits:

Caution

In high-altitude environments, EtCO

2

values may be lower than values observed at sea level, as

described by Dalton’s law of partial pressures. When using the Propaq M unit in high-altitude
environments, it is advisable to adjust EtCO

2

alarm settings accordingly.

Setting Upper FiCO

2

Limits

Initially, the FiCO2 Alarm menu specifies that the FiCO

2

alarms are enabled (

ON

) or disabled

(

OFF

), and displays the default upper limit. The upper limit can be

ON

or

OFF

(default is

OFF

).

The following table lists the default FiCO

2

upper limits for adult, pediatric, and neonate

patients, and gives the range in which you can set these limits:

Patient Type

EtCO

2

Limit Default

EtCO

2

Limit Range

Adult

Lower: 8 mmHg

Upper: 60 mmHg

Lower: 0-145 mmHg

Upper: 5-150 mmHg

Pediatric

Lower: 8 mmHg

Upper: 60 mmHg

Lower: 0-145 mmHg

Upper: 5-150 mmHg

Neonate

Lower: 8 mmHg

Upper: 60 mmHg

Lower: 0-145 mmHg

Upper: 5-150 mmHg

Patient Type

Upper FiCO

2

Limit

Default

Upper FiCO

2

Limit

Range

Adult

8 mmHg

2-98 mmHg

Pediatric

8 mmHg

2-98 mmHg

Neonate

8 mmHg

2-98 mmHg