Invasive pressures – ZOLL Propaq M Rev B User Manual

Page 160

Advertising
background image

Specifications

A-8

www.zoll.com

9650-0803-01 Rev. B

Default Cuff Inflation Pressure:

Adult: 160 mmHg

Pediatric:120 mmHg

Neonatal:90 mmHg

Maximum Cuff Inflation Pressure:

Adult: 270 mmHg

Pediatric:170 mmHg

Neonatal:130 mmHg

Single Fault Backup Overpressure Limit:

Adult: 308 mmHg

Pediatric:205 mmHg

Neonatal:154 mmHg

Typical Determination time without Artifact:

Measurements on the deflation: 30 to 45 seconds

Measurements on the inflation (SureBP)*: 15 to 30 seconds

* using dual lumen cuffs

Maximum Determination Time - Measurement on the Inflation

Adult: 150 seconds

Pediatric:120 seconds

Neonatal:80 seconds

Blood Pressure Validation:

Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method for adults and pediatric patients
and equivalent to an intra-arterial measurement for neonatal patients, within the limits
prescribed by the American National Standards Institute (ANSI-AAMI SP10). To receive a
copy of the report containing the AAMI SP10 results, contact the ZOLL Technical Service
Department.

Invasive Pressures

Number of Channels: 3

Pressure range: -30 to 300 mmHg

Pressure Accuracy: +/- 2 mmHg or 2% of reading, whichever is greater, plus transducer error.

Pulse Rate Range: 25 to 250 BPM

Pulse Rate Accuracy: +/- 3 BPM, or +/- 3% of value whichever is greater

Pulse Rate Display: Average of last 4 beat-to-beat intervals.

Zero Adjust: + / - 200 mmHg