Physiological Alarms (NIBP, SpO2, Resp, CO2, IP & Temp):

Audible: Same as Heart Rate Alarm

Visual: Physiological alarms cause the alarming parameter to be displayed in Red with a white
background. The red device status LED will flash at a rate of 1.7 Hz.

Silence Duration: 90 seconds.

Invalid Operation Alert Tone:

A short, low-pitched tone is audible when a selected control button is unavailable for use or an
invalid entry is detected. Tone frequency is 160 Hz. Duration is 250 msec.

Delays (hold off time):

Heart rate/pulse rate:

i

f source is ECG, 5 seconds

if source is SpO2, 10 seconds

if source is IBP, 5 seconds

if source is NIBP, no hold off

SpO2 Saturation: 10 seconds

EtCO2: 5 seconds

FiCO2: 5 seconds

IBP (Systolic, Diastolic, Mean): 2 seconds

Temperature: 2 seconds

Battery

Type: Rechargeable Lithium-Ion, 11.1Vdc, 6.6 Ah, 73Wh

Capacity:

With a new, fully charged battery operating at room temperature:

At least

7.5 hours of continuous monitoring of ECG, SpO2, CO2, three Invasive Pressure

channels, and 2 channels of Temperature, with NIBP measurements every 15 minutes
(display set to 30%).

Note:

Proper battery care is required to maintain maximum available capacity.

Battery Indicators:

5 Battery capacity LED indicators, Fault indicator, Recalibration indicator

Recharge Rate: 100% in 4 hours, when initiated at Low Battery indication.

General

Weight:

Without Battery: 7.2 lbs.

With Battery: 8.5 lbs.

Dimensions:

Without Handle: 8.9” x 8.7” x 6.5”

With Handle: 8.9” x 10.4” x 7.0”