NIBP System Messages

When monitoring NIBP, the Propaq M unit may display the following messages:

System Message

Cause/Action

READING IN PROGRESS

The unit is taking an NIBP measurement and
functioning normally.

READING STOPPED

The unit has stopped an NIBP measurement,
because the operator has pressed the NIBP
button and cancelled the measurement.

READING FAILED

The patient’s pulse is too weak to for NIBP
measurement, or the cuff requires adjustment.

ARTIFACT

Excessive artifact is preventing the NIBP
measurement.

NEONATE ALERT

The unit has detected a neonate cuff when in
Adult mode. Replace cuff or correct patient
mode, as appropriate.

AIR LEAK

A major air leak is preventing cuff inflation.
Check hose and cuff connections, replace a
defective hose or cuff, as necessary, and
reattempt NIBP measurement.

KINKED HOSE

An air obstruction is preventing the correct
operation of the unit. Check hose for kinks.
Eliminate kinks or replace hose, if necessary.

NIBP DISABLED

A system error has occurred, and the Propaq M
unit should be serviced.

NIBP CALIBRATION CHECK
DUE

The annual calibration check is due. You can
continue to use the device, however, we
recommend that you contact the ZOLL
Technical Service Department or a ZOLL
authorized service representative as soon as
possible to perform the recommended annual
NIBP calibration check.