ZOLL Propaq M Rev B User Manual

Page 162

Advertising
background image

Specifications

A-10

www.zoll.com

9650-0803-01 Rev. B

Electromagnetic Compatibility Guidance and
Manufacturer’s Declaration

In-Flight Use (RTCA/DO-160):

The Propaq M unit complies with RTCA/DO-160, Environmental Conditions and Test
Procedures for Airborne Equipment, using the methods in Section 21, Category M for Radiated
and Conducted Radio Frequency Energy.

General Use Environments (IEC 60601-1-2):

The Propaq M unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the Propaq M should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment – Guidance

RF Emissions CISPR 11

Group 1

The Propaq M unit uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.

RF Emissions CSPR 11

Class A

The Propaq M unit is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.

Harmonic Emission IEC
6100-3-2

Class A

Voltage Fluctuations/ Flicker
Emissions IEC 61000-3-3

Complies

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to EMC information provided in this document.