Monitor/Display

Input: 3-lead, 5-lead, or 12-lead patient cable.

Type: Color LCD, 640 x 480 pixels, 800 MCD

Sweep Speed: 25 mm / sec or 50 mm / sec (User Selectable)

Lead Selections: I, II, III, AVR, AVL, AVF, V1-6.

Frequency Response: (User-selectable)

0.67 to 20 Hz Limited mode

0.67 to 40 Hz Monitor Mode

0.25 to 40 Hz Filtered Diagnostic Mode

0.05 to 150 Hz Diagnostic mode

Per methods a, b, and c of EC11 3.2.7.2.

Common Mode Rejection:

Complies with AAMI EC13-2002 section 4.2.9.10.

Tall T-Wave Rejection:

Meets AAMI EC13-2002, section 4.1.2.1c for 0.9 mV

T-wave (0.8 mV with diagnostic response) and 1mV QRS.

Diagnostic Signals Applied to Patient Connections:

Leads off / active noise suppression sensing circuit is < 0.1 microamps DC. The impedance
/respiration detector signal frequency is 72 ± 7 kHz at 77 microamps RMS
pseudo-sinewave into 100 ohms.

Heart Rate Meter: 30 to 300 BPM.

Displayed Heart Rate: Average of last 5 beat-to-beat intervals.

Heart Rate Alarms: User-selectable.

Size: 0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-ranging.

Heart Rate Meter Response Time:

Responds to a 40 BPM step increase in heart rate within 4.5 seconds per
AAMI EC-13-2002, section 4.1.2.1.f. Responds to a 40 BPM step decrease within
3.9 seconds per AAMI EC-13-2002, section 4.1.2.1.f. Response times include a
1.0-second display update interval.

Heart Rate Response to Irregular Rhythm: (AAMI EC13-2002, section 4.1.2.1.e.)

Ventricular Bigeminy: 80 BPM (expected)

Slow Alternating Ventricular Bigeminy: 60 BPM (expected)

Rapid Alternating Ventricular Bigeminy: 120 BPM (expected)

Bidirectional Systole: 45 BPM (expected)