Using the co2 parameter control panel – ZOLL Propaq M Rev B User Manual

Page 116

Advertising
background image

C

HAPTER

10 M

ONITORING

CO2

10-10

www.zoll.com

9650-0803-01 Rev. B

Setting Upper and Lower Respiratory Rate (RR/BR) Limits

Initially, the RR/BR Alarm menu specifies that RR/BR alarms are enabled (

ON

)

or disabled

(

OFF

), and displays the default Upper and Lower respiratory limits. The upper and lower limits

can be

ON

or

OFF

(default is

OFF

). The following table lists the default Respiratory limits for

adult, pediatric, and neonate patients in Breaths/Minute, and gives the range in which you can
set these limits:

Using the CO2 Parameter Control Panel

Use the navigation keys to highlight and select the CO2 numeric display to display the
CO2 Parameter Control Panel:

Figure 10-2 CO2 Parameter Control Menu

On the CO2 Parameter Control Panel, you can select a CO

2

alarm (EtCO2, Breath Rate, or

FiCO2). On the selected alarm settings menu, you can enable/disable alarms and set alarm
limits (as described previously). You can also set the CO

2

sweep speed.

Patient Type

Respiration Rate Default

Respiration Rate Range

Adult

Lower: 3 BPM

Upper: 50 BPM

Lower: 0 to 145 BPM

Upper: 5 to 150 BPM

Pediatric

Lower: 3 BPM

Upper: 50 BPM

Lower: 0 to 145 BPM

Upper: 5 to 150 BPM

Neonate

Lower: 12 BPM

Upper: 80 BPM

Lower: 0 to 145 BPM

Upper: 5 to 150 BPM

0 to 30 mmHg

0 to 150 mmHg

CO2 Scale

0 to 100 mmHg

0 to 60 mmHg

CO2

Lower

Upper

EtCO2 Alarm

28

50

No Breath Alarm

FiCO2 Alarm

0

10

30 sec

30

RR/BR Alarm

10

CO2/Resp Sweep Speed

12.5 mm/s

Advertising