Applying electrodes to the patient – ZOLL Propaq M Rev B User Manual

Page 67

Advertising
background image

ECG Monitoring Setup

9650-0803-01 Rev. B

Propaq M Operator’s Guide

6-3

Applying Electrodes to the Patient

The following sections show where to place electrodes when using 3- and 5-Lead cables to
perform ECG monitoring. For 3-Lead ECG cables, apply electrodes as in Figure 4-1, 3-Lead
Electrode Placement
. For 5-Lead ECG cables, apply electrodes as in Figure 4-2, 5-Lead
Electrode Placement
.

Avoid placing electrodes over tendons and major muscle masses.

3-Lead Electrode Placement

Depending upon local usage, the ECG leads are marked either RA, LA, and LL (or R, L, and
F). The following table shows the markings and color codes for the different lead sets.

Figure 6-1 3-Lead Electrode Placement

AHA Color Coding

IEC Color Coding

Placement of Electrodes

RA/White Electrode

R/Red Electrode

Place near patient’s right mid-clavicular line,
directly below clavicle.

LA/Black Electrode

L/Yellow Electrode

Place near patient’s left mid-clavicular line,
directly below clavicle.

LL/Red Electrode

F/Green Electrode

Place between 6th and 7th intercostal
space on patient’s left mid-clavicular line.

Advertising